How to solve the problem of abnormal pressure in liquid chromatography?


Release time:

2025-08-26

In the world of analytical chemistry, the Liquid Chromatograph (LC) is a workhorse, meticulously separating complex mixtures into their individual components. Think of it as a sophisticated molecular race track. The mobile phase (liquid) is the track itself, pushing the sample mixture through a column packed with a stationary phase (a special material). The molecules in the sample interact differently with this material, causing them to exit, or "elute," at different times, allowing for precise identification and measurement.

For this molecular race to be fair and consistent, one parameter is king: system pressure. It’s the vital sign of your LC’s health. Just like blood pressure in humans, pressure that is too high, too low, or wildly fluctuating is a clear symptom that something is wrong. Ignoring these warnings can lead to failed experiments, damaged equipment, and costly downtime.

This article will serve as your guide to diagnosing and resolving these common pressure-related ailments, empowering you to keep your LC system running smoothly.

The Goldilocks Zone: What is Normal Pressure?

First, it's crucial to establish a baseline. "Normal" pressure is method-specific. It depends on the column (its length, internal diameter, and particle size), the flow rate, and the viscosity of the mobile phase. A method using a long column packed with tiny particles and a high flow rate will naturally generate higher backpressure than one with a short, wide column and a slow flow rate.

The key is to know your method’s typical operating pressure. Always record this value when the method is first developed and performing well. Any significant deviation from this baseline—be it a sustained increase, a sudden drop, or erratic fluctuations—is your cue to start investigating.

The High-Pressure Headache: Causes and Cures

Sustained high pressure is the most common complaint. It signals an obstruction somewhere in the system, most often after the pump. The increased force is the pump struggling to push liquid through a clog.

Common Culprits and Solutions:

The Column: The Usual Suspect. The column itself is the narrowest point in the entire flow path and is therefore the most likely to get clogged. Contaminants from samples or mobile phases can accumulate on the column frit (the small filter at the end), acting like a kink in a hose.

Solution: Gently flush the column according to the manufacturer's instructions, often starting with a weaker solvent and moving to a stronger one. Using in-line filters and guard columns (a short, sacrificial column placed before the main one) is the best prevention, trapping debris before it reaches the expensive analytical column.

Blocked Frits and Filters: Your LC system has several other frits and filters designed to protect it. The pump inlet filter, the in-line filter between the autosampler and column, and the detector cell filter can all become blocked.

Solution: Methodically check and clean or replace these filters. Start with the easiest-to-access ones before moving to more involved parts.

Tubing Troubles: Over time, especially with high-salt buffers, microcrystalline deposits can form inside the capillary tubing, particularly in the mixing chamber or at connection points, gradually restricting flow.

Solution: Flush the entire system with a strong solvent (like 50/50 water/isopropanol). Check for any tubing that is crimped or bent too tightly, creating a physical blockage.

The Dreaded Pressure Drop: When Force Fades

A sudden or gradual drop in pressure is equally concerning. It typically indicates a failure in the system to maintain flow, almost always due to a leak or an issue with the pump.

Common Culprits and Solutions:

The Leak: Public Enemy #1. LC systems operate under high pressure, so any weakness will become a leak. Common sites include:

Fittings: Ferrule-based connections can loosen over time or from temperature changes.

Seals: The pump seals, which prevent liquid from leaking out around the piston, are wear items and will degrade with use.

The Injector: The rotor seal in the autosampler can wear out from repeated injections.

Solution: Perform a visual inspection. Look for salt crystals or damp areas. Tighten fittings carefully (over-tightening can damage them). For pump seals, consult your manual for replacement schedules and procedures. A tell-tale sign of a leak before the pump (in the solvent lines) is the presence of air bubbles in the lines.

Air Bubbles: The Unwanted Guests. A large bubble trapped in the pump head can cause a massive pressure drop and erratic readings. This is called pump cavitation. The pump is designed to compress liquid, not gas, so the bubble compresses and expands, preventing consistent flow.

Solution: This is a prime cause of pressure fluctuations. Prime the system thoroughly by purging each solvent line individually. For persistent bubbles, a sonication step to degas your solvents can be very effective. Using a dedicated in-line degasser is the best long-term solution.

Failed Check Valves: The pump uses small check valves (inlet and outlet) that act as one-way gates to ensure solvent flows in the correct direction. If a piece of debris prevents one from sealing, or if it wears out, solvent can flow backward, causing low and fluctuating pressure.

Solution: Often, flushing the valves with a strong solvent can dislodge debris. If not, they may need to be replaced.

The Jerky Pulse: Erratic and Fluctuating Pressure

If your pressure trace looks like a seismograph during an earthquake, you’re dealing with fluctuations. This is frequently a combination of the issues above.

Air Bubbles: The most common cause. A small bubble passing through the pump will cause a sharp dip in pressure.

Failing Pump Seal: A worn seal may cause a small, rhythmic leak synchronized with the pump piston stroke, creating a regular "sawtooth" pattern in the pressure readout.

Sticky Check Valve: A valve that is intermittently sticking open or closed will cause sudden, irregular pressure spikes and drops.

Prevention: The Best Medicine

A proactive approach saves immense time and stress.

Use Clean, High-Quality Solvents: Filter and degas your mobile phases.

Prepare Samples Carefully: Centrifuge or filter samples to remove particulate matter.

Embrace Guard Columns: This small investment protects your much more expensive analytical column.

Follow a Maintenance Schedule: Regularly replace seals and purge valves as recommended. Don’t wait for them to fail.

Flush Faithfully: At the end of each day or sequence, flush your system with a weak solvent (e.g., high water content) to remove salts, followed by a strong solvent (e.g., high organic content) to remove hydrophobic contaminants.

Key words:


Contact Information

Add: 2nd Floor, Building 3, No. 1 Chaoqian Rd, Changping, Beijing, China PRC.

HQ: 702, 2#, No.22, Daoan Rd, Suzhou, China.

TEL:86 18613365565

MAIL: info@uvtech-cc.com

Focus on us

erweima

WeChat

erweima

WhatsApp

Copyright 2024 Beijing UVTech Inc.  All Rights Reserved

Business License