Why Must the Mobile Phase Be Filtered and Degassed?


Release time:

2025-06-20

I. Why Filtration is Essential?

Particulates in mobile phases cause:

Column Clogging:

Particles accumulate on inlet frits (0.5-2μm pores), increasing backpressure (>500 psi)

Reduced efficiency: Plate count drops >20% with peak broadening

System Damage:

Pump seal wear: Particles accelerate piston seal degradation

Injector jamming: >10μm particles scratch rotor seals

Baseline Noise: Light scattering in UV flow cells

▌ Filtration Protocol:

Membrane Selection:

Solvent TypeRecommended MembranePore Size
Aqueous/BuffersNylon 660.45μm
Organics (ACN/MeOH)PTFE0.22μm

Apparatus:
Use glass filtration units (organic resistance) at ≤0.3MPa vacuum

Critical Notes:

Filter buffers AFTER salt dissolution (Never add salts post-filtration!)

Change membranes when switching solvents


II. Science Behind Degassing

Dissolved gases (O₂, N₂, CO₂) induce:

Pump Failures:

Bubbles cause pressure pulsation (±10% fluctuation)

Accelerated plunger wear in low-pressure mixing systems

Detector Interference:

UV baseline spikes: Refractive index changes from bubbles

Signal drift in electrochemical detection (O₂ redox reactions)

Retention Time Shifts:
Bubbles occupy ~5% column volume, altering flow paths

▌ Degassing Methods Comparison:

MethodEfficiencyApplicationLimitations
Sonication60-70%Small volumes (<2L)40% regas in 30 min
Helium Sparging>95%Oxygen-sensitive analysesHigh helium cost
Online Degasser98%UHPLC/Gradient analysisMembrane replacement
Vacuum Filtration80%Buffer solutionsVolatile loss

▶ Best Practices:

Staged Degassing:

Step 1: Sonicate 20 min (40kHz, 25°C)

Step 2: Helium sparge 15 min (0.5L/min)

Step 3: Vacuum filter through 0.22μm membrane

Online Degasser Maintenance:

Weekly flush with 0.1M HNO₃ (30 min)

Replace PTFE tubing every 6 months


III. Troubleshooting Degassing-Related Failures

Case 1: Periodic Baseline Fluctuation (>0.5mAU)

Diagnosis: Microbubbles in pump head causing flow instability

Solution:

Purge pump at 5mL/min for 10 min with purge valve open

Increase degasser temp to 35°C (for ACN/water)

Case 2: Retention Time Drift (>±0.2min)

Root Cause: CO₂ dissolution forming carbonic acid (pH shift)

Corrective Actions:

Prepare buffers fresh, store under nitrogen

Add 0.001% triethylamine (suppresses CO₂ absorption)

Key words:


Contact Information

Add: 2nd Floor, Building 3, No. 1 Chaoqian Rd, Changping, Beijing, China PRC.

HQ:1-11 Kanda lzumi cho, Chiyoda ku, Tokyo, Japan

TEL:18613365565

MAIL: info@uvtech-cc.com

Focus on us

erweima

WeChat

erweima

WhatsApp

Copyright 2024 Beijing UVTech Inc.  All Rights Reserved

Business License